Flow Conductivity Sensor
Conductivity measurement is accomplished by using an advanced platinum seven ring quartz cell.
The advantage of this design is that there are no platinum black surfaces which can be contaminated or can deteriorate during profiling or monitoring.
The sensor can be easily cleaned in the field.
| Cell dimensions: | Inner diameter 8 mm, length 45 mm. |
| Cell type: | Seven platinum rings fused inside a quartz tube. |
| Range: | 0..350 mS. |
| Response time: | 50 ms (1). |
| Operating pressure: | 700 bar. |
| Maximum pressure: | 700 bar. |
| Output connections: | 6 x 0.4 mm insulated copper wires (two rings use the same wire). |
| Mounting: | Through 12 mm hole with two 0-ring seals (Parker 2-12). |
| Weight: | 65 gr. |
| Body: | Black plastic and titanium body. |
| Filling for pressure compensation: | Silicone oil. |
(1) At 1 m/second flow rate.

N.B. Dimensions are in millimeters.


N.B. Dimensions are in millimeters.

N.B. Dimensions are in millimeters.
A Crystal IS UVC LED (Ultraviolet 250-300 nm@500µW Light-Emitting Diode) is integrated into the conductivity sensor quartz cell (patent pending). The UVC LED sterilizes the early growth of biofouling, thus eliminating environmental drift in the conductivity sensor.

N.B. Dimensions are in millimeters.
Features

N.B. Dimensions are in millimeters.
| Operating pressure: | 700 bar. |
| Maximum pressure: | 700 bar. |
| Mounting: | Through 12 mm hole with two 0-ring seals (Parker 2-12). |
| Weight: | 50 gr. |
| Body: | Titanium. |
| Filling for pressure compensation: | Silicone oil. |
Conductivity:
| Cell dimensions: | Inner diameter 8 mm, length 45 mm. |
| Cell type: | Seven platinum rings fused inside a quartz tube. |
| Range: | 0..350 mS. |
| Response time: | 50 ms (1). |
| Output connections: | 6 x 0.4 mm insulated copper wires (two rings use the same wire). |
| Body: | Black plastic |
(1) At 1 m/second flow rate.
Temperature:
| Type: | FP07 @ 0 °C thermoresistor. |
| Range: | -30..80 °C. |
| Drift: | Less than 0.001 °C/year. |
| Response time: | 20 ms. |

N.B. Dimensions are in millimeters.