pH sensor

pH SENSOR

Product literature - Brochure

Cylindrical, high pressure glass membrane pH electrode.

Technical specifications:

Membrane type: uranium glass membrane (500Mohm @ 20 °C).
Null point: 7 pH.
Range: 1 .. 13 pH.
   
Maximum pressure: 700 bar
Drift: 0.05 pH/month.
Response time: 3 s.
Operating pressure: 600 bar.
Output connections: shielded wire and LEMO plug.
Mounting: through 12 mm hole with two 0-ring seals (Parker 2-12).
Weight: 35 gr. (including the hydrating cap).
Body: titanium.
 Note: complete with hydrating cap half-filled with pH 7 Buffer Solution.
 
pH SENSOR (Glass) (700 bar)

 

NOTE: Dimensions are in millimeters.