Polarographic sensor, with replaceable cap for membrane support, pressure compensated. Negligible stirring effect, pump is not needed. Technical specifications: | Type: | polarographic sensor with platinum-iridium cathode and silver anode (99.99%). | | Output current: | typical 60 +/- 15 nA in air @ 20o C (green membrane). | | Zero output current: | less than 50 pA, with nitrogen. | | Polarization voltage: | 650 mV DC. | | Range: | 0 .. 50 ppm O2. | | Drift: | 5%/month typical. | | Response time: | 3 s (green membrane @ 20o C). 0.9 s (blue membrane @ 20°C) | | Operating pressure: | 150 bar. | | Output connections: | two 0.4 mm insulated copper wires. | | Mounting: | through 12mm hole with two 0-ring seals (Parker 2-12). | | Weight: | 30 gr. | | Body: | black plastic (PPS). | |
| |
| |
| DISSOLVED OXYGEN SENSOR (150 bar) |
 |
Polarographic sensor, with replaceable cap for membrane support, pressure compensated. Negligible stirring effect, pump is not needed. |
| Type: | polarographic sensor with platinum-iridium cathode and silver anode (99.99%). | | Output current: | typical 60 +/- 15 nA in air @ 20o C (green membrane). | | Zero output current: | less than 50 pA, with nitrogen. | | Polarization voltage: | 650 mV DC. | | Range: | 0 .. 50 ppm O2. | | Drift: | 5%/month typical. | | Response time: | 3 s (green membrane @ 20o C). 0.9 s (blue membrane @ 20°C) | | Operating pressure: | 600 bar. | | Maximum pressure: | 700 bar. | | Output connections: | two 0.4 mm insulated copper wires. | | Mounting: | through 12mm hole with two 0-ring seals (Parker 2-12). | | Weight: | 30 gr. | | Body: | titanium. | | DISSOLVED OXYGEN SENSOR (700 bar) | |