DISSOLVED OXYGEN

DISSOLVED OXYGEN

Product literature - Brochure

Polarographic sensor, with replaceable cap for membrane support, pressure
compensated. Negligible stirring effect, pump is not needed.


Technical specifications:

Type: polarographic sensor with platinum-iridium cathode and
silver anode (99.99%).
Output current: typical 60 +/- 15 nA in air @ 20o C (green membrane).
Zero output current: less than 50 pA, with nitrogen.
Polarization voltage: 650 mV DC.
Range: 0 .. 50 ppm O2.
Drift: 5%/month typical.
Response time: 3 s (green membrane @ 20o C).
0.9 s (blue membrane @ 20°C)
Operating pressure: 150 bar.
Output connections: two 0.4 mm insulated copper wires.
Mounting: through 12mm hole with two 0-ring
seals (Parker 2-12).
Weight: 30 gr.
Body: black plastic (PPS).
 
 
DISSOLVED OXYGEN SENSOR (150 bar)
Polarographic sensor, with replaceable cap for membrane support, pressure
compensated. Negligible stirring effect, pump is not needed.
Type: polarographic sensor with platinum-iridium cathode and
silver anode (99.99%).
Output current: typical 60 +/- 15 nA in air @ 20o C (green membrane).
Zero output current: less than 50 pA, with nitrogen.
Polarization voltage: 650 mV DC.
Range: 0 .. 50 ppm O2.
Drift: 5%/month typical.
Response time: 3 s (green membrane @ 20o C).
0.9 s (blue membrane @ 20°C)
Operating pressure: 600 bar.
Maximum pressure: 700 bar.
Output connections: two 0.4 mm insulated copper wires.
Mounting: through 12mm hole with two 0-ring
seals (Parker 2-12).
Weight: 30 gr.
Body: titanium.
DISSOLVED OXYGEN SENSOR (700 bar)
Polarographic sensor. Negligible stirring effect, pump is not needed
Type: polarographic sensor with platinum-iridium cathode and
silver anode (99.99%).
Output current: typical 60 +/- 15 nA in air @ 20o C.
Zero output current: less than 50 pA, with nitrogen.
Polarization voltage: 650 mV DC.
Range: 0 .. 50 ppm O2.
Drift: 5%/month typical.
Response time: 30 s (membrane @ 20o C).
Operating pressure: 5 bar.
Weight: 30 gr.
Output connections: two 0.4 mm insulated copper wires.
Mounting: through 12mm hole with two 0-ring
seals (Parker 2-12).
Weight: 30 gr.
Body: black plastic (PPS).
MAINTENANCE-FREE DISSOLVED OXYGEN SENSOR (5 bar)
 
DISSOLVED OXYGEN SENSOR WITH ANALOGUE OUTPUT

The sensor consists of a POM-made housing with integrated electronic device, exchangeable sensor head (with possibility of membrane and electrolyte exchange) and an underwater connector.
Measuring principle: amperometric, membrane covered sensor
Housing: plastic (black POM)
Depth range: less than 50 pA, with nitrogen.
Measuring range: 0 ÷ 200 % saturation
Resolution: 0,1 % saturation
Accuracy: 1,5 % saturation
Response time: 1 ÷ 3 sec. (depends on the choice of membrane)
Power supply: 8 ÷ 24 VDC
Signal output: 0 ÷ 5 VDC
Connector: SEACON® XSG-4-BCL
(Mating connector: SEACON® RMG-4-FS)
   
   
DISSOLVED OXYGEN SENSOR WITH ANALOGUE OUTPUT  
 

DISSOLVED OXYGEN SENSOR WITH ANALOGUE OUTPUT
Schematic

NOTE: Dimensions are in millimeters.