Submersible battery pack

Submersible battery pack

Product literature - Brochure

DESCRIPTION
The OCEAN SEVEN submersible 150-bar and 700-bar rechargeable battery packs are housed in a plastic and
an AISI316L container respetively and are composed of 12 NiMH 1.2VDC cells, assembled in series, to create
a unique 14.4VDC 4.5Ah battery pack. The battery pack comes complete with a plastic flange (to join the battery
pack to the OCEAN SEVEN 3xx probes), with an international battery charger and with the submersible cable
needed to connect the battery pack to the Ocean Seven 3xx probes.

Electrical: Voltage 14.4 VDC (nominal voltage).
Capacity 4.5 Ah.
Cells 12 cells – 1.2VDC NiMH type HR-4/3FAU
Charger international battery charger.
Housing: 150 bar 700 bar
Material White plastic POM. AISI 316L.
Diameter 75 mm. 79 mm.
Length 315 mm without top connector. 330 mm without top connector.
Weight 2.1 kg (including batteries). 10 kg (including batteries).
Max pressure 150 bar. 700 bar.
Connector 2-pole SEA CON® connector. 2-pole SEA CON® connector.

INSTRUCTIONS ON HOW TO RECHARGE THE BATTERY PACK

» Dry the battery pack and lay it down on a table.
» Loosen the four closing screws on the cover with the hexagonal
wrench included in the probe accessories.
» Pull the screws completely out and gently move the cover to
one side of the battery pack.
» Dry any trace of water in the proximity of the external side of the
o-ring.

» Connect the battery charger to the top head connector.
» Read the below “Battery Charger” instructions.
» Put the cover on the battery pack.
» Check the correct position of the o-ring and close the battery
pack again with the four screws.
» Gradually tighten the screws alternating one screw with another.

(DO NOT TIGHTEN EXCESSIVELY - PLASTIC POM HOUSING).

» It is enough that you tighten a quarter of a turn of the screw at
the end of stroke of the screw thread (plastic POM housing).

OCEAN SEVEN SUBMERSIBLE
150-BAR AND 700-BAR
RECHARGEABLE BATTERY PACKS
BATTERY CHARGER


Connecting the battery pack, the charger immediately checks the nominal voltage thus automatically selecting
the most appropriate action to carry out as:
1. If the battery voltage is lower than 1.2 V x cell, the charger carries out the standard charging procedure
which implies the recharging at a moderate current for 30m. Whenever the battery reaches the nominal
voltage, the charger automatically starts the quick-charging procedure.
2. If the battery voltage is higher than 1.2 V x cell, the charger carries out the quick-charging procedure.
3. If the battery voltage is lower than 0.2 V x cell, the charger carries out the standard charging procedure
thus inhibiting the quick-charging procedure.

Note
At the end of the quick-charging procedure, the charger undertakes a low-current equilibration charging
procedure which lasts 30 m. This procedure is also called trickily-charging procedure. The aim of this procedure
is to keep the battery cells in good conditions thus compensating for the auto-discharge.

LED INDICATORS

Green ON: charger powered.
Yellow BLINKS: standard charging procedure/equilibration procedure.
Yellow ON: quick charging procedure.
Yellow OFF: batteries are fully charged.
Red ON: bad battery; immediately stop the charging procedure and disconnect the
battery


SPECIFICATIONS

Power: 100 – 250 VAC 50/60Hz, 25 W.
Protection: 2 A fuse, transient voltage suppressor.
Reverse battery polarity.
Reverse current.
Charge current: 1.1 A quick charge, maximum time 6 hours.
120 mA standard charging procedure, maximum time 30m.
120 mA equilibration charging procedure, maximum time 15m.
40 mA trickily-charging procedure
Time-out: 8 hours.
Hold-off: 15 m.
Cells: this charger has been configured for 12 1.2V cells.
PRECAUTIONS


» Do not charge warm/hot batteries.
» Do not charge batteries if the air temperature is higher than +35°C.
» Do not charge lead batteries.
» Do not charge primary batteries like: Alkaline, Mercury, Lithium, etc.. they may EXPLODE.
» Do not charge in the field
» Do not try to recharge a closed battery pack.